[SENSe:]
[SENSe:]FREQuency:STARt
Specifies the start (lowest) frequency for the frequency band of the current measurement.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|CPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
In FFT analysis instrument mode (INST:SEL FFT):
The values of FREQ:STAR and FREQ:SPAN define the frequency bandwidth. The size of the
bandwidth (FREQ:SPAN) remains constant if the start frequency changes.
Allowable start frequency values are defined by the following formulas:
single channel measurements (without option AY6):
0 ≤ start frequency ≤ (115 kHz - (frequency span/2))
two channel measurements or single channel measurements with option AY6:
0 ≤ start frequency ≤ (57.5 kHz - (frequency span/2))
four channel measurements:
0 ≤ start frequency ≤ (28.75 kHz - (frequency span/2))
Step size (FREQ:STEP) determines the change in frequency which results when you send UP or DOWN
with this command.
See figure 18-2 for the range of start frequency values with different bandwidths.
18-46
[SENSe:]FREQuency:STARt {<number>[<unit>]}|<step>|<bound>
limits:
0.0:114999.9023
OUTPUT 711;"SENS:FREQ:STAR 1000"
OUTPUT 711;":freq:star 4.8e3"
[SENSe:]FREQuency:STARt?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0
SCPI Compliance:
no
confirmed
command/query