Display[:Window[1|2|3|4]]:Trace:y[:Scale]:Topcommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Bits in the Standard Event event register are set to 1 under the following conditions:
Operation Complete (bit 0) is set to one when the following two events occur (in the order listed):
You send the *OPC command to the analyzer.
The analyzer completes all pending overlapped commands (see "Command Synchronization"
earlier in this chapter).
Request Control (bit 1) is set to 1 when both of the following conditions are true:
The analyzer is configured as an addressable-only GPIB device (see "Controller Capabilities"
earlier in this chapter).
The analyzer is instructed to do something (such as plotting or printing) that requires it to take
control of the bus.
Query Error (bit 2) is set to 1 when the command parser detects a query error.
Device Dependent Error (bit 3) is set to 1 when the command parser detects a device-dependent error.
Execution Error (bit 4) is set to 1 when the command parser detects an execution error.
Command Error (bit 5) is set to 1 when the command parser detects a command error.
Power On (bit 7) is set to 1 when you turn on the analyzer.
Figure 1-11 also shows the commands you use to read and write the Standard Event registers. See
chapter 20 for more information about these commands.
1-21