[SENSe:]FREQuency:RESolution
Specifies the frequency measurement resolution for FFT and swept sine instrument modes.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|PCT|PNT/SWP|PNT/DEC|PNT/OCT]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
In FFT analysis instrument mode (INST:SEL FFT):
Variable resolution is available and can be set to 100, 200, 400 or 800 lines. Frequency resolution is
unitless. See table 18-3 to determine the number of displayed data points with variable resolution.
This command is similar to FREQuency:BLOCksize which specifies the frequency resolution by
setting the length of the time record.
[SENSe:]FREQuency:RESolution {<number>[<unit>]}|<step>|
<bound>
limits:
0.015625:51200.0
(swept sine only)
OUTPUT 711;"SENS:FREQ:RES 100 PNT/SWP"
OUTPUT 711;":freq:res 800"
[SENSe:]FREQuency:RESolution?
Real
Option:
not applicable
Synchronization Required:
Preset State:
400 (FFT and correlation)
101 PNT/SWP (swept sine)
SCPI Compliance:
no
confirmed
[SENSe:]
command/query
18-31