Common Commands
*TRG
Triggers the analyzer when TRIG:SOUR is BUS.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
This command triggers the analyzer when the following two conditions are met:
The GPIB is designated as the trigger source. (See the TRIG:SOUR BUS command.)
The analyzer is waiting to trigger. (Bit 5 of the Operation Status register must be set). It is ignored at
all other times.
The *TRG command has the same effect as TRIG:IMM. It also has the same effect as the GPIB bus
management command Group Execute Trigger (GET).
3-18
*TRG
OUTPUT 711;"*trg"
OUTPUT 711;"*Trg"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
confirmed
command