CALCulate
CALCulate[1|2|3|4]:UNIT:DBReference:USER:LABel
Assigns a name to the Y-axis unit when CALC:UNIT:DBR 'USER'.
Command Syntax:
<NAME> ::= ASCII characters - 32 through 126
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The name assigned with this command labels the display's Y- axis. A prefix of "dB" is attached to the
name. The label appears only in dB magnitude trace coordinates (CALC:FORM MLOG) and when the
user dB magnitude reference level has been specified with the CALC:UNIT:DBR 'USER' command.
The trace specifier, CALCulate [1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
6-116
CALCulate[1|2|3|4]:UNIT:DBReference:USER:LABel '<NAME>'
maximum number of characters:
OUTPUT 711;"CALC:UNIT:DBR:USER:LAB 'g'"
OUTPUT 711;"calculate2:unit:dbreference:user:label
'm/s^2'"
CALCulate[1|2|3|4]:UNIT:DBReference:USER:LABel?
"STRING"
Option:
not applicable
Synchronization Required:
Preset State:
"V"
SCPI Compliance:
5
yes
instrument-specific