TRIGger
The maximum post-trigger delay cannot exceed 1 Msec (1 x 10
6
seconds). If the specified
post-trigger delay is greater than 1 Msec, the analyzer uses 1 Msec.
The difference between the delay specified for the "earliest" channel and the delay specified for the
"latest" channel is dependent upon frequency resolution (FREQ:RES) as shown in table 24-1. In
correlation and histogram instrument modes, resolution is always 400 lines.
In octave analysis instrument mode:
This command specifies the delay between triggering and the start of an octave measurement.
The maximum post-trigger delay is 99,999 seconds. The minimum is 0 seconds. Only a
post-trigger delay, entered as a positive quantity (+), is valid. A pre-trigger delay is not available
in octave measurements.
The post-trigger delay is applied to all channels. The channel specifier is not used in octave
instrument mode.