Agilent Technologies 35670A Manual page 644

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

TRIGger
TRIGger:TACHometer:HOLDoff
Specifies a time interval in which the tachometer trigger is inhibited.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command causes the tachometer to ignore inputs for the specified time after each accepted
tachometer event. It "holds off" the trigger for the tachometer for a specified amount of time.
See the ARM:SOURce commands for additional information about RPM step arming. See the
[SENSe:]ORD commands for additional information about order tracking.
24-12
TRIGger:TACHometer:HOLDoff {<number>[<unit>]}|<step>|
<bound>
limits:
0.0:0.052224
OUTPUT 711;"Trigger:tachometer:holdoff 0.05s"
OUTPUT 711;"trig:tach:hold 2"
TRIGger:TACHometer:HOLDoff?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0
SCPI Compliance:
no
instrument-specific
command/query

Advertisement

Table of Contents
loading

Table of Contents