[SENSe:]SWEep:DWELl
Specifies the integration time for swept sine measurements.
Command Syntax:
<dwell_time> ::= {<number>[<unit>]}|<step>|<bound>
<number> ::= a real number (NRf data)
<unit> ::= [S|CYCLE]
<step ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Integration time is the amount of time that each point is measured.
Sending SWE:DWEL in seconds, results in a constant integration scale.
Sending SWE:DWEL in cycles, results in a proportional integration scale. At higher frequencies the
same number of cycles occurs in a shorter time. The integrate time must be a minimum of 1 cycle long.
The analyzer takes any value less than one as one complete cycle.
[SENSe:]SWEEP <dwell_time>
limits:
250e-6:32768 S
or
1:234 CYCLE
OUTPUT 711;"sense:sweep:dwel .005 s"
OUTPUT 711;"Swe:Dwell 1 cycle"
[SENSe:]SWEep:DWELl?
Real
Option:
1D2 Swept Sine
Synchronization Required:
Preset State:
+5.0 CYCLE
SCPI Compliance:
no