Agilent Technologies 35670A Manual page 300

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay
DISPlay[:WINDow[1|2|3|4]]:WATerfall:HEIGht
Specifies the height of a waterfall trace box.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [PCT]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The value you specify determines the height of the waterfall trace box as a percentage of the total height
of the waterfall display area. See figure 8-2.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
8-46
DISPlay[:WINDow[1|2|3|4]]:WATerfall:HEIGht {<num-
ber>[<unit>]}|<step>|<bound>
limits:
1:100
OUTPUT 711;"Disp:Window4:Wat:Height 69"
OUTPUT 711;"DISP:WAT:HEIGHT 84"
DISPlay[:WINDow[1|2|3|4]]:WATerfall:HEIGht?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
+57 PCT
SCPI Compliance:
Figure 8-2. Height of the Waterfall Trace Box
as a Percentage of the Total Height
of the Waterfall Display Area
no
instrument-specific
command/query

Advertisement

Table of Contents
loading

Table of Contents