Calculate[1|2|3|4]:Synthesis:gaincommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:CFIT:FREQuency:STOP
Specifies the stop frequency for a curve fit operation over a limited frequency span.
Command Syntax:
<number> ::= a real number (NRf data
<unit> ::= [HZ]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is not valid unless the CALC:CFIT:FREQ OFF command has been sent.
Note
This command is not trace specific. It ignores the trace specifier.
CALCulate[1|2|3|4]:CFIT:FREQuency:STOP {<num-
ber>[<unit>]}|<step>|<bound>
limits:
0.390625:115000.0
OUTPUT 711;":Calculate:Cfit:Freq:Stop 1.2 khz"
OUTPUT 711;"CALC:CFIT:FREQ:STOP 800"
CALCulate[1|2|3|4]:CFIT:FREQuency:STOP?
Real
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
+5.12E+004
SCPI Compliance:
no
instrument-specific
CALCulate
command/query
6-11