Agilent Technologies 35670A Manual page 350

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

INPut
INPut[1|2|3|4]:BIAS[:STATe]
Enables/disables the ICP supply on the corresponding input channel.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command connects (or disconnects) the internal 4 mA current source to the input connector. The
nominal voltage output is 24 V dc (open circuit).
To get rid of the DC bias that is generated when INP:BIAS is ON, turn on AC coupling. See the
INPut:COUPling command.
If the channel specifier is not used, the command defaults to channel 1.
12-2
INPut[1|2|3|4]:BIAS[:STATe] OFF|0|ON|1
OUTPUT 711;"INPUT3:BIAS:STATE ON"
OUTPUT 711;"inp3:bias OFF"
INPut[1|2|3|4]:BIAS[:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
confirmed
command/query