Agilent Technologies 35670A Manual page 442

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]AVERage:HOLD
Specifies the type of hold used in averaging octave measurements.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the type of average hold used for octave measurements. It is valid when used
with:
Linear averaging (AVER:TYPE RMS).
Exponential averaging (AVER:TYPE RMS;TCON EXP).
Equal confidence averaging (AVER:TYPE ECON).
If you send AVER:HOLD MAX, the analyzer displays the maximum averaged spectrum value for each
band.
If you send AVER:HOLD MIN, the analyzer displays the minimum averaged spectrum value for each
band. This is useful for estimating background noise.
Note
AVERage:HOLD differs from the peak hold function. See the AVER:TYPE MAX
command description for information about the peak hold function.
18-4
[SENSe:]AVERage:HOLD OFF|0|MAXimum|MINimum
OUTPUT 711;"sens:average:hold MINIMUM"
OUTPUT 711;"Aver:Hold OFF"
[SENSe:]AVERage:HOLD?
OFF|MAX|MIN
Option:
1D1 Realtime Octave
Synchronization Required:
Preset State:
OFF
SCPI Compliance:
no
instrument-specific
command/query