Agilent Technologies 35670A Manual page 452

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]AVERage:TCONtrol
Specifies how the analyzer behaves after the count (AVER:COUN) is reached.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies termination control during averaging.
To specify linear (normal) averaging, send TCON FRE with the AVER:TYPE command.
To specify exponential averaging, send TCON EXP with the AVER:TYPE command.
To specify repeat averaging ("autostart"), send TCON REP with the AVER:TYPE command. The
analyzer takes N averages, clears the data, waits for arming conditions, and then takes another N
averages. The analyzer continues taking measurements until you send one of the following commands:
INIT:CONT OFF
AVER:STAT OFF
AVER:TCON FRE
Depending upon the instrument mode (specified with the INST:SEL command), some types of
termination control and averaging are not valid. See table 18-1.
18-14
[SENSe:]AVERage:TCONtrol FREeze|REPeat|EXPonential
OUTPUT 711;"Sense:Aver:Tcon FREEZE"
OUTPUT 711;"AVERAGE:TCON REPEAT"
[SENSe:]AVERage:TCONtrol?
FRE|REP|EXP
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
confirmed
command/query