MMEMory
MMEMory:STORe:TCAPture
Saves the time capture buffer to a file on the specified disk.
Command Syntax:
<FILENAME> ::= '[<PATH>]<filename>'
<PATH> ::= <MSIS>[{DOS_DIR}]
<MSIS> ::= RAM:|NVRAM:|INT:|
<DOS_DIR> ::= \<DIR_NAME>\
<DIR_NAME> ::= ASCII characters
<filename> ::= ASCII characters (see beginning of chapter
Example Statements:
Attribute Summary:
Description:
This command saves the current time capture buffer to a file on the specified disk.
If the buffer is too large for the disk, an error, "Media full; File too Large" is generated. Use the
MMEM:STORE CONT command to split the file over multiple disks. See the MMEM:STORE CONT
command for more information about splitting a time capture file on multiple disks.
If the filename you specify matches the name of another file on the disk, this command overwrites the old
file.
Note
If you plan to transfer this file to a PC, refer to the MMEM:STOR:TRAC:FORM
command.
15-36
MEMory:STORe:TCAPture <FILENAME>
EXT[,<select_code>[,<unit _number>]]:
(DOS only)
for <filename> restrictions)
OUTPUT 711;"mmem:stor:tcap, 'int:tcap1'"