[SENSe:]ORDer:TRACk[1|2|3|4|5]:STATe
Selects order track or order spectrum measurements.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The analyzer makes an order track measurement when ORD:TRAC:STAT ON is sent.
The analyzer makes an order spectrum measurement when ORD:TRAC:STAT OFF is sent.
Note
This command ignores the order track specifier. If the state is changed for one order
track, the state changes for all order tracks.
[SENSe:]ORDer:TRACk[1|2|3|4|5]:STATe OFF|0|ON|1
OUTPUT 711;":ORD:TRAC2:STATE OFF"
OUTPUT 711;"sens:order:trac2:stat OFF"
[SENSe:]ORDer:TRACk[1|2|3|4|5]:STATe?
Integer
Option:
1D0
Computed Order Tracking
Synchronization Required:
Preset State:
+0 (all tracks)
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query