CALCulate[1|2|3|4]:MARKer[:STATe]
Turns on the main markers or turns off all markers and marker functions for a selected trace.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To display the main marker and its annotation, send CALC:MARK ON. The analyzer displays the X-axis
and Y-axis values at the top of the grid.
To disable the display of the main markers and the marker reference for the active trace, send
CALC:MARK OFF.
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
Note
All markers and marker functions operate on the specified trace, whether or not the trace
is displayed.
CALCulate[1|2|3|4]:MARKer[:STATe] OFF|0|ON|1
OUTPUT 711;":CALCULATE3:MARK OFF"
OUTPUT 711;"calculate:mark:stat OFF"
CALCulate[1|2|3|4]:MARKer[:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
ON (+1)
SCPI Compliance:
no
instrument-specific
CALCulate
command/query