TRIGger:EXTernal:LEVel
Specifies the level of the external input signal which causes the analyzer to trigger.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [V]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Use this command to trigger the analyzer when the signal applied to the external trigger input connector
passes through the specified value in the direction set by the TRIG:SLOP command. The external trigger
connector is on the rear panel of the analyzer.
When TRIG:EXT:RANGE is set to HIGH, the allowable range is -10 V to +10 V. The level may be set
to 78.125 mV increments. When TRIG:EXT:RANGE is set to LOW, the allowable range is -2V to +2V.
The level may be set to 15.625 mV increments.
This command is only valid with external triggering (TRIG:SOUR EXT).
TRIGger:EXTernal:LEVel {<number>[<unit>]}|<step>|<bound>
limits:
-10.0:10.0
OUTPUT 711;":Trigger:Ext:Level -0.412237"
OUTPUT 711;"TRIG:EXT:LEVEL 0.54171"
TRIGger:EXTernal:LEVel?
Real
Option:
not applicable
Synchronization Required:
Preset State:
0 V
SCPI Compliance:
no
instrument-specific
TRIGger
command/query
24-3