MMEMory
MMEMory:STORe:LIMit:UPPer:TRACe[1|2|3|4]
Saves the upper limit of the specified trace to a file on the specified disk.
Command Syntax:
<FILENAME> ::= '[<PATH>]<filename>'
<PATH> ::= <MSIS>[{DOS_DIR}]
<MSIS> ::= RAM:|NVRAM:|INT:|
<DOS_DIR> ::= \<DIR_NAME>\
<DIR_NAME> ::= ASCII characters
<filename> ::= ASCII characters (see beginning of chapter
Example Statements:
Attribute Summary:
Description:
This command saves an upper limit to a file. The trace specifier selects which upper limit you are
saving—TRAC1 for trace A, TRAC2 for trace B, TRAC3 for trace C, or TRAC4 for trace D. Trace A is
the default if you do not specify a trace.
If the file name you specify matches the name of another file on the disk, this command overwrites the
old file.
Note
If you plan to transfer this file to a PC, refer to the Standard Data Format Utilities Users
Guide .
15-30
MMEMory:STORe:LIMit:UPPer:TRACe[1|2|3|4] <FILENAME>
EXT[,<select_code>[,<unit _number>]]:
(DOS only)
for <filename> restrictions)
OUTPUT 711:mmem:store:limit:upper:trace2 'int:newlim'"
OUTPUT 711;"MMEM:STOR::LIM:UPP:TRACE1, 'LIMIT1.LIM'"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command