Agilent Technologies 35670A Manual page 444

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]AVERage:IRESult:RATE
Specifies how often the display is updated when fast average mode is on.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the rate used by the analyzer when AVER:IRES is on.
The analyzer updates the display once for each N averages. N is the update rate specified with this
command. The analyzer continues to update the display whenever it reaches a multiple of N.
18-6
[SENSe:]AVERage:IRESult:RATE <number>|<step>|<bound>
limits:
1:9999999
OUTPUT 711;"Sens:Average:Ires:Rate 8575803"
OUTPUT 711;"AVERAGE:IRES:RATE 3430970"
[SENSe:]AVERage:IRESult:RATE?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
+5
SCPI Compliance:
no
instrument-specific