Agilent Technologies 35670A Manual page 262

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay
DISPlay:GPIB:ECHO
Enables and disables the echoing of GPIB command mnemonics to the analyzer's screen.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When echoing is enabled, the analyzer displays the GPIB command mnemonic which corresponds to the
operation executed from the front panel. The command mnemonic appears on the third line in the
upper-left corner of the screen.
Not every keystroke generates an GPIB command.
8-8
DISPlay:GPIB:ECHO OFF|0|ON|1
OUTPUT 711;":Disp:Gpib:Echo OFF"
OUTPUT 711;"DISPLAY:GPIB:ECHO ON"
DISPlay:GPIB:ECHO?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
command/query