SOURce
SOURce:VOLTage[:LEVel]:REFerence
Specifies the amplitude of the reference input channel for the autolevel feature in swept sine
measurements (INST:SEL SINE).
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [DBVRMS|VPK|DBVPK|V|DBV|EU|DBEU|VRMS]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the amplitude which the analyzer tries to maintain for the input reference channel
when autolevel is enabled (SOUR:VOLT:AUTO ON).
19-12
SOURce:VOLTage[:LEVel]:REFerence {<number>[<unit>]}|
<step>|<bound>
limits:
-69.276:31.66
OUTPUT 711;"SOUR:VOLT:LEV:REF:CHAN .1 VPK"
OUTPUT 711;"sour:volt:ref:chan 2 VRMS"
SOURce:VOLTage[:LEVel]:REFerence?
Real
Option:
1D2 Swept Sine
Synchronization Required:
Preset State:
+2.0 VPK
SCPI Compliance:
no