CALCulate
CALCulate[1|2|3|4]:CFIT:DATA
Loads values into the curve fit table.
Command Syntax:
<BLOCK> ::= see Description
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command transfers a complete curve fit table from your controller to the analyzer.
When you transfer a curve fit table to the analyzer, you must use the definite length block syntax. Data
must be 64-bit binary floating-point numbers (see the FORMat[:DATA] REAL command). The elements
of the definite length block for a curve fit table are defined below.
The first point specifies the table type; followed by three points which specify the size of the table. Th
next 80 points are the complex pairs. The remaining points are descriptors. Points 152 - 174 indicate
which terms are fixed. Points 173 - 175 are adjusters; gain, frequency scale and time delay.
See "Block Data" in the GPIB Programmer's Guide for additional information about transferring block
data.
Note
This command is not trace specific. It ignores the trace specifier.
6-6
CALCulate:CFIT:DATA <BLOCK>
OUTPUT 711;":CALCULATE:CFIT:DATA BLOCK"
OUTPUT 711;"calc:cfit:data BLOCK"
CALCulate[1|2|3|4]:CFIT:DATA?
DEF_BLOCK
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command/query