CALCulate[1|2|3|4]:MARKer:MAXimum[:GLOBal]
Moves the main marker to the highest point on the specified trace.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
This command moves the marker to the highest peak one time. Another command
—CALC:MARK:MAX:TRAC—controls a marker function that automatically moves the marker to the
highest peak each time the trace is updated.
The specified trace does not need to be displayed for the marker to move to the highest peak.
CALCulate[1|2|3|4]:MARKer:MAXimum[:GLOBal]
OUTPUT 711;":Calc:Mark:Maximum"
OUTPUT 711;"CALC:MARKER:MAX:GLOB"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
command
6-75