Agilent Technologies 35670A Manual page 445

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]AVERage:IRESult[:STATe]
Selects fast average mode.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies whether the analyzer displays in fast average mode.
In FFT and correlation instrument mode, AVER:IRES ON updates the display once for every N averages.
N is the update rate specified with AVER:IRES:RATE. The preset AVER:IRES rate is 5. OFF updates
the display after each average.
In swept sine instrument mode, AVER:IRES ON updates the display after the entire sweep is completed.
OFF updates the display at each point in the sweep.
In histogram instrument mode, AVER:IRES ON updates the display as fast as it can without slowing
down the measurement. OFF updates the display at the end of the measurement.
[SENSe:]AVERage:IRESult[:STATe] OFF|0|ON|1
OUTPUT 711;":aver:ires ON"
OUTPUT 711;"Sense:Aver:Iresult:Stat OFF"
[SENSe:]AVERage:IRESult[:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query
18-7

Advertisement

Table of Contents
loading

Table of Contents