HCOPy
HCOPy:PAGE:DIMensions:USER:LLEFt
Specifies the lower left position (P1) of the plot area.
Command Syntax:
<P1_X-axis_value> ::= <number>|<step>|<bound>
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
<P1_Y-axis_value> ::= <number>|<step>|<bound>
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The plot area is defined by the X-axis and Y-axis values for the scaling points P1 and P2. See figure
10-2. The X-axis value and the Y-axis value depends upon the plotter and the number of points on the
page.
Use the HCOP:PAGE:DIM:USER:URIG command to specify the P2 X- and Y-axis values.
Check the documentation for your plotter to select appropriate P1 values.
Note
HCOP:PAGE:DIM:USER:LLEF is set to +332,+1195 after a reset (*RST).
10-20
HCOPy:PAGE:DIMensions:USER:LLEFt {<P1_X-axis_value>,
<P1_Y-axis_value>}
limits:
-32767:32767
limits: -32767:32767
OUTPUT 711;"HCOP:PAGE:DIM:USER:LLEF 332,4286"
OUTPUT 711;"hcopy:page:dimensions:user:lleft 4743,1195"
HCOPy:PAGE:DIMensions:USER:LLEFt?
Integer, Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
no
instrument-specific
command/query