[SENSe:]
[SENSe:]AVERage:PREView
Specifies the type of preview averaging.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is valid in FFT analysis mode only (INST:SEL FFT).
To enable manual previewing, send AVER:PREV MAN. The analyzer waits for a response before taking
the next time record.
To enable timed previewing, send AVER:PREV TIM. The analyzer waits for a specified amount of time
before accepting the time record. That is, if no response is sent the analyzer accepts the time record. Set
the time period with the AVER:PREV:TIME command.
After each time record is collected, the Waiting for Accept/Reject bit in the Operation Status Register is
set to 1. The bit is cleared when the analyzer receives an accept or reject command or when the analyzer
receives a command that changes the measurement setup.
To accept the time record send AVER:PREV:ACC. To reject the time record send AVER:PREV:REJ.
To turn off average previewing, send AVER:PREV OFF.
18-8
[SENSe:]AVERage:PREView OFF|0|MANual|TIMed
OUTPUT 711;"SENS:AVERAGE:PREV TIMED"
OUTPUT 711;"average:prev TIMED"
[SENSe:]AVERage:PREView?
OFF|MAN|TIM
Option:
not applicable
Synchronization Required:
Preset State:
OFF
SCPI Compliance:
no
instrument-specific
command/query