[SENSe:]AVERage:IMPulse
Enables impulse detection in octave measurements.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When impulse detection is on (AVER:IMP ON), the analyzer computes and displays the IEC 651 impulse
characteristics in the overall power band.
In linear averaging (AVER:TYPE RMS;TCON FRE), the analyzer calculates the value of the impulse
output over the average time (specified with the AVER:TIME command).
In all other types of averaging, the analyzer calculates the instantaneous value of the impulse vector.
See the analyzer's Online Help for additional information.
[SENSe:]AVERage:IMPulse OFF|0|ON|1
OUTPUT 711;":AVER:IMPULSE ON"
OUTPUT 711;"sens:aver:impulse ON"
[SENSe:]AVERage:IMPulse?
Integer
Option:
1D1 Realtime Octave
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query