SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet
Specifies a DC offset for the source output.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [V]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Use this command to set a DC bias on the analyzer's source.
The allowable range is -10 V to + 10 V. If the DC offset is between -2 V and +2 V, it can be set in 1 mV
increments; otherwise it is set in 5 mV increments.
The analyzer resets the DC offset level to the nearest valid increment if the the sum of the source level
(set with the SOURce:VOLTage[:LEVel][:IMMediate][:AMPlitude] command) and the DC offset exceed
±10 Volts.
SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet {<number>
[<unit>]}|<step>|<bound>
limits:
-10.0:10.0
OUTPUT 711;"SOUR:VOLTAGE:LEV:IMM:OFFSET 9.40924"
OUTPUT 711;"sour:voltage:offs -2.92258"
SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0 V
SCPI Compliance:
no
confirmed
SOURce
command/query