Calculate[1|2|3|4]:Math:selectcommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate
CALCulate[1|2|3|4]:CFIT:ABORt
Aborts the curve fit operation.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
This command aborts the current curve fit operation. As a result of the CALC:CFIT:ABOR command,
the curve fit table is not updated.
Note
This command is not trace specific. It ignores the trace specifier.
6-4
CALCulate[1|2|3|4]:CFIT:ABORt
OUTPUT 711;":Calc3:Cfit:Abort"
OUTPUT 711;"CALC:CFIT:ABOR"
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command