Agilent Technologies 35670A Manual page 468

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]FREQuency:MANual
Selects a discrete point to be measured during manual sweep mode.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The frequency must fall within the start and stop frequencies (FREQ:STAR and FREQ:STOP). Multiple
points can be measured by repeating this command with each value.
Note
This command is only used if the SWE:MODE MAN command has been sent.
18-30
[SENSe:]FREQuency:MANual {<num-
ber>[<unit>]}|<step>|<bound>
limits:
0.015625:51200.0
OUTPUT 711;":FREQUENCY:MAN 1388.13"
OUTPUT 711;"sense:freq:man 24158.9"
[SENSe:]FREQuency:MANual?
Real
Option:
1D2
Swept Sine
Synchronization Required:
Preset State:
51.2 HZ
SCPI Compliance:
no
confirmed
command/query