CALCulate[1|2|3|4]:MARKer:MAXimum:LEFT
Moves the main marker one peak to the left of its current location on the specified trace.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
A peak is a local maximum on the displayed trace. The slope of a trace is positive to the left of a peak
and negative to the right. In addition, the slope on one side of a peak must not change for at least one
vertical division (one-third division on both sides if in octave analysis instrument mode or one-tenth
division (one decade) if using logarithmic scaling of the Y-axis).
This command only finds peaks that are at least one point to the left of the current marker position. If the
analyzer does not find a peak, the marker does not move. You can increase the number of peaks found by
the analyzer by decreasing the value of vertical scale division (DISP:TRAC:Y:PDIV).
CALCulate[1|2|3|4]:MARKer:MAXimum:LEFT
OUTPUT 711;":CALC3:MARKER:MAX:LEFT"
OUTPUT 711;"calc:mark:maximum:left"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
command
6-77