CALCulate
CALCulate[1|2|3|4]:MARKer:Y[:ABSolute]?
Returns the main marker's Y-axis position.
Query Syntax:
Example Statements:
Return Format:
Attribute Summary:
Description:
This query always returns the Y-axis position of the main marker, even if the marker is not currently
displayed on the analyzer's screen. The returned value tells you the amplitude of the specified trace at the
marker's X-axis position (specified with CALC:MARK:X or CALC:MARK:POS:POIN).
Send CALC:MARK:Y? UNIT to determine the units for the Y-axis.
6-92
CALCulate[1|2|3|4]:MARKer:Y[:ABSolute]?
OUTPUT 711;"CALCULATE4:MARK:Y:ABS?"
OUTPUT 711;"calc3:marker:y?"
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
query