Agilent Technologies 35670A Manual page 503

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]SWEep:SPACing
Selects linear or logarithmic spacing between measurement data points.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Send SWE:SPAC LIN for linearly spaced frequency points. The frequency step size does not change
during the sweep of the frequency points; it remains constant over the entire spectrum.
Send SWE:SPAC LOG for logarithmically or proportionately spaced frequency points. The ratio of the
location of adjacent points is constant. In addition to hertz, decade or octave units may be used with the
FREQ:SPAN command. The following units may be used with the FREQ:RES command:
PNT/SWP
PNT/DEC
PNT/OCT
PCT (spacing between measurement points as a percentage of the total frequency span)
[SENSe:]SWEep:SPACing LINear|LOGarithmic
OUTPUT 711;":SWE:SPAC LINEAR"
OUTPUT 711;"sense:swe:spacing LOGARITHMIC"
[SENSe:]SWEep:SPACing?
LIN|LOG
Option:
1D2
Swept Sine
Synchronization Required:
Preset State:
LIN
SCPI Compliance:
no
confirmed