Agilent Technologies 35670A Manual page 289

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay[:WINDow[1|2|3|4]]:TRACe:Y[:SCALe]:AUTO
Scales and repositions the trace vertically to provide the best display of trace data.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To initiate autoscaling of the specified trace, send DISP:TRAC:Y:AUTO ONCE. The analyzer's
autoscaling algorithm changes the values of DISP:TRAC:Y:REF and DISP:TRAC:Y:PDIV to optimize
Y-axis scaling which provides the best display of your data.
ON autoscales the specified trace after every display update. OFF disables autoscaling.
The trace specifier, WINDow [1|2|3|4], determines which trace you are selecting. Omit the specifier or
send WIND1 for trace A, WIND2 for B, WIND3 for C, or WIND4 for D.
DISPlay[:WINDow[1|2|3|4]]:TRACe:Y[:SCALe]:AUTO OFF|0|ON|
1|ONCE
OUTPUT 711;"DISPLAY:WIND2:TRAC:Y:SCAL:AUTO ONCE"
OUTPUT 711;"disp:trac:y:auto ON"
DISPlay[:WINDow[1|2|3|4]]:TRACe:Y[:SCALe]:AUTO?
ON|OFF
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
confirmed
DISPlay
command/query
8-35

Advertisement

Table of Contents
loading