Calculate[1|2|3|4]:Unit:x:user:frequency:factorcommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:FORMat
Selects a coordinate system for displaying measurement data and for transferring coordinate transformed
data to a controller.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To select a coordinate system that displays linear magnitude data along the Y-axis, send CALC:FORM
MLIN.
To display linear magnitude data on a logarithmic Y-axis scale, send CALC:FORM
MLIN;:DISP:TRAC:Y:SPAC LOG.
To select a coordinate system that displays logarithmic magnitude data on a linear Y-axis scale, send
CALC:FORM MLOG.
To select a coordinate system that displays wrapped phase along the Y-axis, send CALC:FORM PHAS.
Phase wraps at the display boundaries. If you change the display boundaries with the
DISP:WIND:TRAC:Y:TOP and DISP:WIND:TRAC:Y:BOTTOM commands, you change where phase
wraps on the display.
To select a coordinate system that displays imaginary numbers along the Y-axis, send CALC:FORM
IMAG. This coordinate system shows the imaginary component of complex data at each point along the
X-axis. If the data point is real rather than complex, a value of 0 is displayed for all X-axis points.
To select a coordinate system that displays real numbers along the Y-axis, send CALC:FORM REAL.
This coordinate system shows real data or the real component of complex data at each point along the
X-axis. This and the polar diagram (CALC:FORM POLar) are the only valid selection for orbit displays
(CALC:FEED 'XVOL:VOLT 1,2'). CALC:FORM REAL is the default selection for orbit displays.
CALCulate[1|2|3|4]:FORMat
MLINear|MLOGarithmic|PHASe|REAL|IMAGinary|NYQuist|UPHase
|GDELay|POLar
OUTPUT 711;"CALC3:FORMAT IMAGINARY"
OUTPUT 711;"calc4:format IMAGINARY"
CALCulate[1|2|3|4]:FORMat?
MLIN|MLOG|PHAS|REAL|IMAG|NYQ|UPH|GDEL|POL
Option:
not applicable
Synchronization Required:
Preset State:
MLOG trace A and trace C
REAL trace B and trace D
SCPI Compliance:
yes
confirmed
CALCulate
command/query
6-29

Advertisement

Table of Contents
loading