[SENSe:]FREQuency:BLOCksize
Specifies the number of real-time data points displayed on the analyzer's screen.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command sets the number of points in a time record for FFT and correlation instrument modes.
In FFT analysis instrument mode (INST:SEL FFT) the number of frequency-data points depends upon the
blocksize. See table 18-2. This command is similar to the FREQuency:RESolution command which
specifies the resolution with frequency lines.
In correlation analysis instrument mode (INST:SEL CORR) the number of displayed time-data points
depends upon the blocksize. See table 18-2.
[SENSe:]FREQuency:BLOCksize <number>|<step>|<bound>
limits:
256:2048
OUTPUT 711;":FREQUENCY:BLOCK 512"
OUTPUT 711;"sens:freq:bloc 1024"
[SENSe:]FREQuency:BLOCksize?
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query
18-27