Agilent Technologies 35670A Manual page 274

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay
DISPlay[:WINDow[1|2|3|4]]:LIMit:STATe
Turns limit lines on or off in the specified display.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Sending DISP:LIM:STAT ON only enables the display of the limit lines in the specified trace. For
example, send DISP:WIND3:LIM:STAT ON to turn on the limit line for trace C. To test the trace against
those lines, you must send CALC3:LIM ON. If a trace specifier is not used, the command defaults to
trace A.
Note
A trace can be evaluated against limits even when limit lines are not displayed.
8-20
DISPlay[:WINDow[1|2|3|4]]:LIMit:STATe OFF|0|ON|1
OUTPUT 711;"disp:window:lim:stat OFF"
OUTPUT 711;"Display:Lim:State ON"
DISPlay[:WINDow[1|2|3|4]]:LIMit:STATe?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
command/query

Advertisement

Table of Contents
loading

Table of Contents