Calculate[1|2|3|4]:Marker[:State]Command/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

ARM
ARM:RPM:MODE
Enables the Start RPM Arming qualifier.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
For a runup measurement send ARM:RPM:MODE UP. The first arm occurs when the RPM value
reaches the specified threshold value (specified with ARM:RPM:THR).
For a rundown measurement send ARM:RPM:MODE DOWN. The first arm occurs when the RPM
value reaches the specified threshold value (specified with ARM:RPM:THR).
To specify subsequent arms to occur at RPM intervals measured from the starting RPM value see the
ARM:TIMer command.
In order analysis instrument mode (INST:SEL ORD; Option 1D1), you can disable the Start RPM
Arming qualifier. Send ARM:RPM:MODE OFF.
5-4
ARM:RPM:MODE OFF|0|UP|DOWN
OUTPUT 711;"Arm:Rpm:Mode UP"
OUTPUT 711;"ARM:RPM:MODE DOWN"
ARM:RPM:MODE?
OFF|UP|DOWN
Option:
not applicable
Synchronization Required:
Preset State:
UP
SCPI Compliance:
no
instrument-specific
command/query