INITiate
INITiate:CONTinuous
Sets the trigger system to a continuously initiated state.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
INIT:CONT OFF suspends the measurement process. It "pauses" the current measurement immediately.
The analyzer discards the time record that is currently being processed.
INIT:CONT ON restarts a "paused" measurement. It also allows you to add more data to the running
average of a completed measurement. For example, if the analyzer has completed a 10-average
measurement and you send INIT CONT ON, 10 more records are averaged with the old data, bringing the
total number of averages to 20.
INIT:CONT ON is not valid if you pause a measurement and change the measurement parameters. It is
also not valid, if a calibration occurs while a measurement is paused. You must use the INIT[:IMM]
command to start a paused measurement after calibration or after changing the measurement parameters.
Note
After *RST, INIT:CONT is set to OFF (+0).
11-2
INITiate:CONTinuous OFF|0|ON|1
OUTPUT 711;"Init:Cont OFF"
OUTPUT 711;"INITIATE:CONT OFF"
INITiate:CONTinuous?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
ON (+1)
SCPI Compliance:
no
confirmed
command/query