ARM
ARM:SOURce
Specifies the type of arming for the analyzer's trigger.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To select automatic arming send IMM. The analyzer waits for the hardware to settle and then waits for a
trigger signal (specified by TRIG:SOUR) before starting the measurement. When the measurement is
completed, the trigger is automatically re-armed.
To select manual arming send MAN. The analyzer waits for the hardware to settle then waits for the
ARM[:IMM] command, and then waits for a trigger signal before starting the measurement. The
ARM[:IMM] command must be sent to re-arm the trigger after the measurement is completed.
To select RPM step arming send RPM. The analyzer waits for the hardware to settle, waits for the next
RPM level to be reached, and then waits for the trigger signal before starting the measurement. See the
TRIGger:TACHometer commands for more information about using the analyzer's tachometer.
To select time step arming send TIM. The instrument waits for the hardware to settle, waits for a trigger
and starts the measurement by collecting the first time record. The analyzer then waits until the end of
the time step interval (specified by ARM:TIM) and waits for the trigger signal before collecting the next
time record. In waterfall displays, all waterfall traces are referenced from the start time. See figure 5-2.
5-6
ARM:SOURce IMMediate|MANual|RPM|TIMer
OUTPUT 711;":arm:sour MANUAL"
OUTPUT 711;"Arm:Sour MANUAL"
ARM:SOURce?
IMM|MAN|RPM|TIM
Option:
not applicable
Synchronization Required:
Preset State:
IMM
SCPI Compliance:
no
confirmed
command/query