Agilent Technologies 35670A Manual page 472

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]FREQuency:RESolution:AUTO
Selects auto resolution for swept sine instrument mode.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The frequency spacing between measurement points is adjusted automatically by the analyzer. The
analyzer increments or decrements the size of the step to accommodate varying frequency response
changes.
The analyzer calculates the ratio of the frequency response of the current point to the frequency response
of the previous point. If the ratio exceeds the maximum percentage change specified by the
FREQ:RES:AUTO:MCH command, the analyzer adjusts the resolution to measure the next point.
The adjusted resolution value is never less than the minimum resolution value specified by the
FREQ:RES:AUTO:MIN command. The initial resolution of the sweep between the first two points is
specified by FREQ:RES:AUTO:MIN.
Note
FREQ:RES:AUTO is ON after a reset (*RST).
See table 18-5 to determine the number of displayed data points with swept sine variable resolution.
18-34
[SENSe:]FREQuency:RESolution:AUTO OFF|0|ON|1
OUTPUT 711;"Sense:Freq:Resolution:Auto OFF"
OUTPUT 711;"FREQ:RESOLUTION:AUTO OFF"
[SENSe:]FREQuency:RESolution:AUTO?
Integer
Option:
1D2
Swept Sine
Synchronization Required:
Preset State:
+0
SCPI Compliance:
no
confirmed
command/query