[SENSe:]
[SENSe:]TCAPture:STOP[1|2|3|4]
Specifies the end of the time capture data used in a measurement.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|BLK|PNT]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command allows you to select a portion of the time capture data to be used for the measurement.
You can set the "analysis region" for each channel independently.
Use the SENS:TCAP:STARt command to specify the beginning of the time capture data.
The analyzer sets this value to the end of the capture data upon receiving any of the following commands:
MMEMory:LOAD:STATe
MMEMory:LOAD:TCAPture
[SENSe:]TCAPture[:IMMediate]
SYSTem:PRESet
*RST
18-76
[SENSe:]TCAPture:STOP[1|2|3|4] {<number>[<unit>]}|<step>|
<bound>
limits:
-9.9e37:9.9e37
OUTPUT 711;":tcap:stop3 10 blk"
OUTPUT 711;"sense:tcapture:stop 4 s"
[SENSe:]TCAPture:STOP[1|2|3|4]?
Real
Option:
not applicable
Synchronization Required:
Preset State:
end of data
SCPI Compliance:
no
instrument-specific
command/query