Common Commands
*PCB
Sets the pass-control-back address.
Command Syntax:
<primary_address> ::= <number>|<step>|<bound>
<secondary_address> ::= <number>|<step>|<bound>
Example Statements:
Attribute Summary:
Description:
Use this command to specify the address of your controller before you pass control of the GPIB to the
analyzer. When the analyzer completes the operation that required it to have control of the bus, it
automatically passes control back to the controller at the specified address.
The optional second number is only used for controllers that support extended addressing. It is ignored
by the analyzer.
The address last specified with this command is saved in nonvolatile memory, so it is not affected when
you turn the analyzer off and on. It is also not affected by the *RST command.
3-10
*PCB <primary_address>, [<secondary_address>]
OUTPUT 711;"*PCB 11"
OUTPUT 711;"*PCB 19"
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
no
confirmed
command