Agilent Technologies 35670A Manual page 276

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay
DISPlay[:WINDow[1|2|3|4]]:POLar:ROTation
Specifies the angle of rotation for the polar diagram.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command rotates the X-axis of a polar diagram. The setting of the DISP:POLar:CLOCkwise
command affects the parameter of this command. See figure 8-1.
If DISP:POLar:CLOCkwise is OFF, a positive value rotates the polar diagram up and to the left of the
horizontal axis and a negative value rotates the polar diagram down and to the left.
If DISP:POLar:CLOCKwise is ON, a positive value rotates the polar diagram down and to the left of the
horizontal axis and a negative value rotates the polar diagram up and to the left.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
8-22
DISPlay[:WINDow[1|2|3|4]]:POLar:ROTation <num-
ber>|<step>|<bound>
limits:
-360:+360
OUTPUT 711;"Display:Wind3:Polar:Rot -311"
OUTPUT 711;"DISP:POLAR:ROT -76"
DISPlay[:WINDow[1|2|3|4]]:POLar:ROTation?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
0 ° - right side of a horizontal X-axis
SCPI Compliance:
no
instrument-specific
command/query

Advertisement

Table of Contents
loading