Agilent Technologies 35670A Manual page 623

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Description:
This command copies a selected trace into one of eight data registers, copies data between data registers,
and replaces existing data in the specified data register with block data from the controller.
The first parameter specifies the destination. The second parameter specifies the data or the source of the
data:
one of the trace boxes; A (TRAC1), B (TRAC2), C (TRAC3), or D (TRAC4)
a calibration trace for one of the input channels
data from one of the data registers
See table 23-1 for the types of measurement data.
*
Baseband measurements have a start frequency of 0 Hz; zoom measurements have a start
frequency
0 Hz.
This command differs from the MMEM:STOR:TRAC command. The MMEM:STOR:TRAC command
saves trace data to a file. TRAC:DATA saves trace data to one of the data registers.
TRACe
*
*
23-3