Agilent Technologies 35670A Manual page 535

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

SOURce:FREQuency[:CW]
Sets the frequency of the sine source.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command sets the frequency of the fixed sine source type. To select a fixed sine source output, use
the SOUR:FUNC SIN command.
The allowable range is 0 to 115 kHz. The frequency may be set to 15.625 mHz increments.
This is an alias for the SCPI command SOURce:FREQuency:FIXed.
SOURce:FREQuency[:CW] {<number>[<unit>]}|<step>|<bound>
limits:
0.0:115000.0
OUTPUT 711;"sour:freq 1 khz"
OUTPUT 711;"SOURCE:FREQUENCY:CW 12000"
SOURce:FREQuency[:CW]?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+1.024E+004
SCPI Compliance:
no
confirmed
SOURce
command/query
19-3

Advertisement

Table of Contents
loading