CALCulate[1|2|3|4]:UNIT:ANGLe
Specifies the unit for phase coordinates.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is only valid when phase trace coordinates are specified (CALC:FORM PHAS or
CALC:FORM UPH).
To select phase units in degrees for the specified trace, send CALC:UNIT:ANGL DEGR. To select phase
units in radians for the specified trace, send CALC:UNIT:ANGL RAD.
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
CALCulate[1|2|3|4]:UNIT:ANGLe DEGRee|RADian
OUTPUT 711;"CALC:UNIT:ANGL DEGR"
OUTPUT 711;"calculate2:unit:angl rad"
CALCulate[1|2|3|4]:UNIT:ANGLe?
DEGR|RAD
Option:
not applicable
Synchronization Required:
Preset State:
DEGR
SCPI Compliance:
yes
confirmed