Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

INPut[1|2|3|4][:STATe]
Specifies one-channel, two-channel or four-channel measurements.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To select a two-channel measurement, send INP2 ON. With Option AY6, send INP2 ON;:INP4 OFF.
To select a one-channel measurement, send INP2 OFF. The analyzer takes data from Channel 1 only.
To select a four-channel measurement (only available with Option AY6), send INP4 ON.
Note
INP1 OFF is not a valid command. The Channel 1 input can not be disabled.
INPut[1|2|3|4][:STATe] OFF|0|ON|1
OUTPUT 711;":inp2 OFF"
OUTPUT 711;"Input:Stat ON"
INPut[1|2|3|4][:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
Standard
INPut1 ON (+1)
INPut2 ON (+1)
INPut3 not available
INPut4 not available
SCPI Compliance:
no
confirmed
INPut
command/query
Option AY6
INPut1 ON (+1)
INPut2 ON (+1)
INPut3 OFF (+0)
INPut4 OFF (+0)
12-9