INPut
INPut[1|2|3|4]:LOW
Sets the specified channel's input shield to float or to ground.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To connect the analyzer's input shield to ground through 55 Ω, send INP[1|2|3|4]:LOW GRO.
To float the analyzer's input shield through 1 MΩ, send INP[1|2|3|4]:LOW FLO. The input connector
ground is not completely isolated from the chassis ground.
If the channel specifier is not used, the command defaults to channel 1.
12-6
INPut[1|2|3|4]:LOW GROund|FLOat
OUTPUT 711;"Input3:Low GROUND"
OUTPUT 711;"INP2:LOW FLOAT"
INPut[1|2|3|4]:LOW?
GRO|FLO
Option:
not applicable
Synchronization Required:
Preset State:
FLO
SCPI Compliance:
no
confirmed
command/query