Agilent Technologies 35670A Manual page 526

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]WINDow[1|2|3|4]:FORCe
Specifies the width of the force window.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the length of the force window in seconds.
The force window passes the first part of the time record (specified by the length of the width of the force
window) and sets the remaining part to the average value of the time record's remaining data. The data is
then multiplied by the exponential windowing function.
The channel specifier, WINDow[1|2|3|4], is ignored. The force width is applied to all channels.
Note
The first point in the time record is always considered to be time t=0.
18-88
[SENSe:]WINDow[1|2|3|4]:FORCe {<number>[<unit>]}|<step>|
<bound>
limits:
3.8147E-6:9.9999E6
OUTPUT 711;":WINDOW:FORCE 0.1 s"
OUTPUT 711;"SENS:WIND:FORC 2 ms"
[SENSe:]WINDow[1|2|3|4]:FORCe?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+9.999E+003 S
SCPI Compliance:
no
confirmed
command/query